To charge the battery, connect the USB charging
cable as shown.
The device cannot be used for any measuremen
during charging.
Use USB charging devices comply with electrica
safety standard, for example IEC 60950.
6.2 Care and Cleaning
Clean the device by carefully swabbing the device surface with a soft cloth swab
with water or alcohol.
6.3 Trouble Shooting
Problem
The device does
not turn on.
The ECG waveform
amplitude is small
ECG waveform
drifts
SpO
or pulse rate
2
shows no value, or
the number
fluctuates
"Error XX"
occurred.
SpO
value is too
2
low when
measured using
integrated sensor.
Have the device repaired by authorized service centers only, otherwise its warranty
is invalid.
Possible Cause
1. The battery may be low.
2. The device might be
damaged
The lead you choose is not
suitable for you.
1. The pressure exerted on
the electrode is not stable or
too much.
2. Hand or body may be
moving.
1. Finger may not be
insertedcorrectly.
2. Finger or hand may be
moving.
Software or hardware
failure.
1. Finger pressed too hard.
2. Finger may not be
insertedcorrectly.
12
Solution
1. Charge the battery and
try again.
2. Please contact with your
local distributor.
Change another lead and
try again.
1. Hold the device stably
and gently.
2. Try to keep perfectly still
and test again.
1. Remove finger and
reinsert, as directed.
2. Try to keep perfectlystill
and test again.
Restart the device. If the
error persists, contact with
authorized service center.
1. Reinsert your finger
gently andstably.
2. Make sure your fingeris
in right position.