Input type
RTD
Lin.R
EMC immunity influence .....................................................
Effect of supply voltage variation ......................... ≤ 0,005% of span / VDC
Vibration ......................................................................... IEC 60068-2-6 Test FC
Lloyd's specification no. 1 ........................................ 4 g / 2...100 Hz
Max. wire size ................................................................ 1 x 1.5 mm
Humidity ......................................................................... < 95% RH (non-cond.)
Dimensions .................................................................... Ø 44 x 20.2 mm
Tightness (enclosure / terminal) ............................ IP68 / IP00
Weight ............................................................................. 50 g
Electrical specifications, input:
RTD and linear resistance input:
RTD
type
Pt100
Ni100
Lin.R
Max. offset ...................................................................... 50% of selec. max. value
Cable resistance per wire (max.) ............................ 10 Ω
Sensor current .............................................................. > 0.2 mA, < 0.4 mA
Effect of sensor cable resistance
(3-wire) ............................................................................ < 0.002 Ω/Ω
Sensor error detection .............................................. Yes
Output:
Current output:
Signal range .................................................................. 4...20 mA
Min. signal range ......................................................... 16 mA
updating time .............................................................. 135 ms
Load resistance ............................................................ ≤ (V
Load stability ................................................................. < ±0.01% of span / 100 Ω
Sensor error detection:
Programmable .............................................................. 3.5...23 mA
18
Basic values
accuracy
≤ ±0.3°C
≤ ±0.2 Ω
Min.
Max.
value
value
-200°C
+850°C
-60°C
+250°C
0 Ω
10000 Ω
Basic
Min.
span
25°C
25°C
30 Ω
supply
Temperature
coefficient
≤ ±0.01°C / °C
≤ ±20 m Ω/ °C
≤
±0.5% of span
2
stranded wire
Standard
IEC 60751
DIN 43760
-----
- 8) / 0.023 [Ω]
IC.PS.P40.A3.ML/520B2284