To increase pressure
Turn the PRV adjusting nut (3) anticlockwise.
To decrease pressure
Turn the PRV adjusting nut (3) clockwise.
6.
After setting the PRV, hold the adjusting nut (3) in position and tighten
the locking nut (2).
Note: The locking nut has a left-hand thread.
7.
Check the operation of the PRV. If necessary reset the PRV.
8.
Secure the locking nut by tightening the two locking screws (1).

Fault Finding

If the system pressure reduces, check the following:
1.
Check the selector valve
tightened.
2.
Allow sufficient time after pressure generation for the temperature to
stabilize. The larger the system volume, the longer the time for
thermal stability.
3.
Check the adaptors, flexible pipe and connections between the pump
and equipment under test for leaks. Tighten any loose joints and
replace any seals that are worn or damaged.
4.
In hydraulic mode, if the volume adjuster can be wound fully in, but
maximum pressure cannot be achieved, there is probably air trapped
in the system.
5.
Check PRV pressure setting.
If, for any reason, a fault occurs within the pump, it is recommended that the
equipment be returned to an appointed agent.
GE Druck PV 411A User Manual
(7) is in the correct position and properly
Re-prime, and repeat.