Beispiel:
Messumformer Typ 2, 1400-0.001mbar
(vgl. Abschnitt 4.2.3)
Istwert:
0.5 mbar
Dekaden: 6
exp0:
-3 (f. 1.0e-3mbar)
V
=
out
(10V / 6) x ( log(0.5) - (-3) ) =
4.50 V ;
3.1.4 Ventil-Anschlüsse
2x Phoenix Combicon, 3pol
vac
vac
24 VDC auf Masse
+ shield
vent
vent
vac: für Vakuumventil
vent: für Belüftungsventil
Example:
transducer type 2, 1400-0.001mbar (s.
chapter 4.2.3)
actual pressure:
0.5mbar
decades:
6
exp0:
-3 (f. 1.0e-3mbar)
V
=
out
(10V / 6) x ( log(0.5) - (-3) ) =
4.50 V ;

3.1.4 Valve Connectors

2x Phoenix Combicon, 3pole
vac
vac
24 VDC to ground
+ shield
vent
vent
vac: for vacuum valve
vent: for venting valve
Exemple:
Convertisseur de mesure type 2,
1400-0.001mbar (p. 4.2.3)
pression réelle:
0.5mbar
décades:
6
exp0:
-3 (f. 1.0e-3mbar)
V
=
out
(10V / 6) x ( log(0.5) - (-3) ) =
4.50 V ;

3.1.4 Raccords de vannes

vac
vac
2x Phoenix Combicon,
3 pôles
24 VDC à terre
vent
vent
+ shield
vac: pour vanne à vide
vent: pour vanne d'échappement