Beurer BF 520 Mode D'emploi page 17

Masquer les pouces Voir aussi pour BF 520:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Table 4
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Recommended separation distances between portable and mobile RF communications equipment
and the BF 520
The BF 520 is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the BF 520 can help prevent electromagnetic interference by main-
taining a minimum distance between portable and mobile RF communications equipment (transmitters) and
the BF 520 as recommended below, according to the maximum output power of the communications equip-
ment.
Rated maximum output power
of transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d in meters (m) can be determined using the equation applicable to the frequency of the transmitter, where
P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufac-
turer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorp-
tion and reflection from structures, objects and people.
Separation distance according to frequency of transmitter
80 MHz to 800 MHz
d =
0.117
0.368
1.167
3.689
11.667
17
m
800 MHz to 2.5 GHz
d =
0.233
0.737
2.333
7.37
23.333

Publicité