Check And Calibration; Important - geo-FENNEL S-Digit 60WL+ Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

6

CHECK AND CALIBRATION

IMPORTANT

Calibration of inclinometer:
- before first use
- before important measurements
- after a hit or drop
- after a large fluctuation in temperature
CHECK OF CALIBRATION
Set the instrument down on a flat and even surface and switch it on. Wait 3 sec.
and note down the measured value (A). Turn the unit by 180° and set it down on
the same position as before. Wait 3 sec. and nte down the measured value (B).
If the deviation between the two values A and B is more than 0,2° the unit has
to be calibrated.
CALIBRATION
Set the instrument down on the same flat and even surface, switch it on and
keep the button CAL / REF pressed until the display shows „-0-" . Wait approx.
3 sec. and press the button CAL / REF again until the display shows „-1-" . Turn
the unit by 180° and set it down on the same position as before. Wait approx.
3 sec. and press the button CAL / REF again until the display shows „-2-" . Now
the calibration is completed.
8
7
9
10
6)
7)
8)
9)
10)
11)
11
Direction symbol (approching
the 0°/90° position)
Symbol sound on/off
Battery status indicator
Unit symbol N/FT
Unit symbol °/%
Unit symbol mm/m
EN
E
13

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

S-digit 120wl+

Table des Matières