CHECK AND CALIBRATION
IMPORTANT
Calibration of inclinometer:
- before first use
- before important measurements
- after a hit or drop
- after a large fluctuation in temperature
CHECK OF CALIBRATION
Set the instrument down on a flat and even surface and switch it on. Wait 3 sec.
and note down the measured value (A). Turn the unit by 180° and set it down on
the same position as before. Wait 3 sec. and nte down the measured value (B).
If the deviation between the two values A and B is more than 0,2° the unit has
to be calibrated.
CALIBRATION
Set the instrument down on the same flat and even surface, switch it on and
keep the button CAL / REF pressed until the display shows „-0-" . Wait approx.
3 sec. and press the button CAL / REF again until the display shows „-1-" . Turn
the unit by 180° and set it down on the same position as before. Wait approx.
3 sec. and press the button CAL / REF again until the display shows „-2-" . Now
the calibration is completed.