Conversion to the
standard hydrogen
electrode potential
T 814246
04/2014
U
= U
+ U
H
Meas
Ref
where:
U
= ORP voltage, relative to the
H
U
= measured ORP voltage
Meas
U
= voltage of the reference system relative to the
Ref
U
is temperature dependent and can be taken from the following table
Ref
(also refer to DIN 38404-6):
T (°C)
U
Ref
TruLine ORP 51
0
+221
5
+216
10
+212
15
+207
20
+202
25
+197
30
+192
standard hydrogen electrode
standard hydrogen electrode
T (°C)
[mV]
35
40
45
50
55
60
Commissioning, measuring, testing
U
[mV]
Ref
TruLine ORP 51
+187
+181
+176
+171
+165
+160