Accessories
Maximum capacity
in kg
50
100
200
500
1)
The gap width of the overload stop may vary dependent on the installation situation. Check the
function of the overload stop after installation and before commissioning. A gap width of 0.05mm
should be present for a load cell loaded with the maximum capacity.
34
Setting the gap width of the overload stop
The screw length of the overload stop is designed for the
use of a ZEL or ZPL. With an optimal gap width, suffi
cient screw-in depth into the base plate (u10 mm) is
ensured. Depending on the load application parts used,
you may need to select different screw lengths, e. g. for
Z6/...KG/ZK M10x35 (DIN 931).
► Set the gap width of the overload stop using a feeler
gage.
► Fix the height setting by locking the screw with the
supplied nut.
1)
Gap A
(overload stop) in mm
A1027-4.1 HBM: public
[0.35
[0.40
[0.50
[0.85
Limit load
200 kg
400 kg
800 kg
2.5 t
Z6...