TECHNICAL DATA

GUIDANCE AND MANUFACTURER'S
DECLARATION ELECTROMAGNETIC EMISSIONS
The AGU NS19 nasal aspirator is intended for use in the electromagnetic envi-
ronment specified below. The customer or the user of the AGU NS19 nasal aspi-
rator should assure that it is used in such an environment.
Emission test
RF emissions
CISPR 11
RF emissions CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
GUIDANCE AND MANUFACTURER'S
DECLARATION ELECTROMAGNETIC IMMUNITY
The AGU NS19 nasal aspirator is intended for use in the electromagnetic environ-
ment specified below.
The customer or the user of the AGU NS19 nasal aspirator should assure that it is
used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
EN
Compliance
Group 1
Class B
Not
applicable
Not
applicable
IEC 60601
Compliance
test level
level
+ 6 kV contact
+ 6 kV contact
+ 8 kV air
+ 8 kV air
3 A/m
3 A/m
Electromagnetic environment guidance
The AGU NS19 nasal aspirator uses RF ener-
gy only for its internal function. Therefore, its
RF emissions are very low and are not likely
to cause any interference in nearby electronic
equipment
The AGU NS19 nasal aspirator is suitable for
use in all establishments, including domestic
establishments and those directly connected
to the public low-voltage power supply
network that supplies buildings used for
domestic purposes
Electromagnetic
environment guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commer-
cial or hospital environment