FW400 Series Blood/Fluid Warmer
5.2
Cleaning
CAUTION
!
Do not immerse Warmer in cleaning and/or
sterilization solution. Do not submerge or soak unit;
it is fluid-resistant, not fluid-proof.
Perform the following care and cleaning between patient use
or per hospital protocol:
Unplug unit prior to cleaning.
To clean the external surfaces of the Warmer, use a
nonabrasive cleaning solution (such as warm soapy water)
and a clean cloth.
To clean the inside of the Warmer, flush the cassette slot with
cleaning solution over a sink. (The slot is open at the bottom
to facilitate cleaning.) Use the thin brush supplied with the
Warmer to clean the inside of slot. Rinse slot with warm
water.
Wipe Warmer dry with a clean, dry cloth.
Apply a disinfectant such as 10% chlorinated bleach
solution (chlorinated bleach with 5.25% sodium
hypochlorite) to the external surfaces of the Warmer and
allow to dry.
5.3
Function Check
A function check should be performed every three (3)
months or as per hospital protocol by qualified medical
service personnel in accordance with the Medi-Temp II
FW400 Series Blood/Fluid Warmer Service Manual (contact
your local dealer or Gaymar to obtain the Service Manual).
9