Cleaning and Disinfecting
Caution
Use only recommended cleaners or disinfectants on surfaces. Immersion-type
disinfectants are not approved for use on surfaces.
Clean and disinfect
The exterior surface of the mini-dock can be cleaned and disinfected with a recommended cleaner or
disinfectant. For recommended disinfectants, see
1 Remove the system, unplug the power supply, and detach any cables.
2 Clean the surfaces using a soft cloth lightly dampened in a mild soap or detergent cleaning solution.
Apply the solution to the cloth rather than the surface.
3 Wipe the surfaces with the disinfecting solution.
4 Air dry or towel dry with a clean cloth.
Compatible cleaners and disinfectants
The following cleaners and disinfectants are recommended for use on the mini-dock.
Table 1-2: Cleaners and disinfectants compatible with the SonoSite mini-dock