5
Technical data
5.2 Error codes
Light absorption too great
Reason e. g.: optical sytem dirty or blank sample too strongly coloured
Check blank sample / measurement compartment.
Adjustment factor "out of range"
Too much light reaching the detector
Reason e. g.: lid not attached
Attach lid.
Battery capacity was too low during measurement
Change battery.
Factory adjustment for U.1 incorrect / erase
User adjustment for U.1 incorrect / erase
Factory adjustment for U.2 incorrect / erase
User adjustment for U.2 incorrect / erase
Factory adjustment for U.3 incorrect / erase
User adjustment for U.3 incorrect / erase
Factory adjustment for U.4 incorrect / erase
User adjustment for U.4 incorrect / erase
Release 03/2009
®
Spectroquant
Colorimeter Picco COD / CSB
E 001
E 010 / E 011
E 020 / E 021
en
E 022
E 070
E 071
E 072
E 073
E 074
E 075
E 076
E 077
31