10. How To Use The Protective Rubber Holster - PEWA BENNING MM 3 Notice D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 20
-
Turn the rotary switch  to the "OFF" position.
-
Remove the protective rubber holster  from the BENNING MM 3.
-
Place the BENNING MM 3 on its front side and remove the three screws
from the case back.
-
Lift the end of the case back near the input terminals until it gently unsnaps
from the case front at the end nearest to the LCD display .
-
Remove the empty battery from inside the case front and carefully
disconnect it from the battery connector leads.
-
Snap the battery connector leads to the terminals of a new battery and
reinsert the battery into the case front. Ensure that the battery leads do not
become pinched between the case back and case front.
-
Press the case back and case front together again and reinstall the three
screws.
-
Place the BENNING MM 3 back into its protective rubber holster .
see Figure 11:
Please contribute to environmental protection! Batteries should
not be thrown into domestic refuse bins! They can be discarded
at collection points for old batteries or special refuse. Please
contact your municipality for more information.
9.4 Fuse replacement
Remove test leads and turn the power off before opening the
BENNING MM 3! Dangerous voltage!
The BENNING MM 3 is protected from overloading by a built-in 1 A (G-cartridge)
fast blow fuse and a built-in 16 A (G-cartridge) fast blow fuse (see figure 12).
This is how to replace the fuses:
-
Remove the safety test leads from the measured circuit.
-
Remove the safety test leads from the BENNING MM 3.
-
Turn the rotary switch  to the "OFF" position.
-
Remove the protective rubber holster  from the BENNING MM 3.
-
Place the BENNING MM 3 on its front side and remove the three screws
from the case back.
-
Lift the end of the case back near the input terminals until it gently unsnaps
from the case front at the end nearest to the LCD display .
Do not remove any screws from the printed circuit board of the
BENNING MM 3!
-
Lift the printed circuit board from the front case.
-
Gently pry one end of the defective fuse loose from the fuse holder.
-
Then slide the defective fuse completely out of the fuse holder.
-
Install a new fuse of the same rating, fast acting characteristic and size.
-
Make sure the new fuse is centered in the fuse holder.
-
Replace the printed circuit board into the case front.
-
Ensure that the battery leads do not become pinched between the case
back and case front.
-
Press the case back and case front together again and reinstall the three
screws.
-
Place the BENNING MM 3 back into its protective rubber holster  again.
see Figure 12:
9.5 Calibration
To maintain the given accuracy of the measurement readings the instrument
must be regularly recalibrated by our service department. We recommend a
calibration interval of one year.

10. How to use the protective rubber holster

-
The safety test leads can be stored by wrapping them around the protective
rubber holster  and then clipping the probes into the protective probe
holders on the rear of the holster  (see figure 13).
-
A safety test lead can be clipped into the probe holder on the protective
rubber holster  with the test probe protruding, in order to apply the probe
together with the BENNING MM 3 to a measuring point .
-
The rear tilt stand on the protective rubber holster  allows the
10/ 2004
Battery replacement
Fuse replacement
BENNING MM 3
17

Publicité

Table des Matières
loading

Table des Matières