 Use shielded, low‐capacitance measurement cables only (HBM cables
fulfill both conditions)
 Do not route the measurement cables parallel to power lines and control
circuits. If this is not possible, protect the measurement cable with steel
conduits, for example
 Avoid stray fields from transformers, motors and contact switches
 Do not ground the transducer, amplifier and indicator more than once.
 Connect all devices in the measuring chain to the same protective earth
conductor
To ensure the best EMC protection, the load cell, together with the connection
cable and the subsequent electronics, should be placed in a shielded housing.