Caractéristiques; Characteristics 2.2. Características - GETINGE ALM/Angenieux ENERGIX WPS Notice Technique

Table des Matières

Publicité

Alimentation pour éclairage opératoire
Power supply for surgical light
Alimentación alumbrado de quirófano
2.2.Caractéristiques
2.2.1. Détail des modules (Figure 2)
WPS 362 / WPS 361 Version A
et B
• Mécaniques
• Thermiques
• Electriques
• Electriques
V
AC+DC
rms
V
AC+DC
rms
• Electriques
2.2.2. Détail carte secours WPS 24
Notice technique/Technical manual/Manual técnico
2.2.Characteristics
2.2.1. Details of WPS 362/WPS 361
modules - Version A and B
(figure 2)
• Mechanical
- aluminium profile, epoxy resin
- dimensions: 120 x 70 x 240
- attachment of options possible on
three faces
• Thermal
- Operating range: 10° to 40°C
- Efficiency: > 85%
- Maximum temperature of radiator:
80°C
- Vertical mounting essential above 200W
- Minimum aluminium panel 350x 300x3
imperative above 200 W
-
MOS technology
• Electrical (primary)
-
Medical equipment class 1 type B
-
Supply voltage
WPS 362: 220/230V ± 10%
WPS 361: 100/120V ± 10%
-
Maximum delivered power 400W
(360W bulbs)
-
Maximum power consumed: 440W
-
Frequency: 50Hz/60Hz
-
Primary protection by fuse: T8A
-
Power factor: 0.92
-
Primary/secondary
> 5000 VDC
-
Leakage current (phase/ground):
< 1 mA (maxi 5 mA)
-
MTBF: 500, 000 hours 30°C ambient
-
Applicable standards:
IEC 601/1/UL 2601
IEC 601/1/2/FCC part 15
-
CE marking according to directive 93/
336 (class 1)
• Electrical (secondary)
- Full wave rectified voltage twice input
frequency
- Chopping frequency: 80Khz
V
- Maximum of adjustable voltage: from
AC+DC
rms
V
24
AC+DC
rms
at power supply output
- Minimum voltage derived from the
maximum voltage: -8 volts (16
)
- Electronic protection against short-
circuits with automatic resetting.
- Slow start 0.2 seconds from power-on
- Output voltage regulation: +/- 1%
over the range covered by the primary
at constant load.
- Maximum module/lighting distance
30m with wire of cross-section 6mm
2
• Electrical (controls)
-
Potentiometer/switch : 10Kohms, travel
W
300°
-
Diode power supply voltage: 2 VDC
-
Switch to batteries indicator (if WPS 24
option fitted), by red diode
-
Maximum control/module distance:
50m with wire of cross-section 0.6mm
2
2.2.2. Details of WPS 24 backup card
• Computation time <0.5s
• Dimension 130 x 55
• Protection by F20A automobile type fuse•
Max power: 400W
• Supply voltage 26VDC ± 4VDC
insulation:
V
to 28
AC+DC
rms
V
AC+DC
rms
Ref. 060123001
ENERGIX WPS
2.2.Características
2.2.1. Detalle de los módulos (Figura 2)
WPS 362 / WPS 361 - Versión A
y B
• Mecánicas
• Térmicas
• Eléctricas
• Eléctricas
V
AC+DC
rms
V
• Eléctricas
2.2.2. Detalle de la tarjeta emergencia
WPS 24
V
AC+DC
rms
AC+DC
rms
2
W
2
9