2.2.Caractéristiques
2.2.1. Détail des modules (Figure 2)
WPS 362 / WPS 361 Version A
et B
Mécaniques
Thermiques
Electriques
Electriques
V
AC+DC
rms
V
AC+DC
rms
Electriques
2.2.2. Détail carte secours WPS 24
Notice technique/Technical manual/Manual técnico
2.2.Characteristics
2.2.1. Details of WPS 362/WPS 361
modules - Version A and B
(figure 2)
Mechanical
- aluminium profile, epoxy resin
- dimensions: 120 x 70 x 240
- attachment of options possible on
three faces
Thermal
- Operating range: 10° to 40°C
- Efficiency: > 85%
- Maximum temperature of radiator:
80°C
- Vertical mounting essential above 200W
- Minimum aluminium panel 350x 300x3
imperative above 200 W
-
MOS technology
Electrical (primary)
-
Medical equipment class 1 type B
-
Supply voltage
WPS 362: 220/230V ± 10%
WPS 361: 100/120V ± 10%
-
Maximum delivered power 400W
(360W bulbs)
-
Maximum power consumed: 440W
-
Frequency: 50Hz/60Hz
-
Primary protection by fuse: T8A
-
Power factor: 0.92
-
Primary/secondary
> 5000 VDC
-
Leakage current (phase/ground):
< 1 mA (maxi 5 mA)
-
MTBF: 500, 000 hours 30°C ambient
-
Applicable standards:
IEC 601/1/UL 2601
IEC 601/1/2/FCC part 15
-
CE marking according to directive 93/
336 (class 1)
Electrical (secondary)
- Full wave rectified voltage twice input
frequency
- Chopping frequency: 80Khz
V
- Maximum of adjustable voltage: from
AC+DC
rms
V
24
AC+DC
rms
at power supply output
- Minimum voltage derived from the
maximum voltage: -8 volts (16
)
- Electronic protection against short-
circuits with automatic resetting.
- Slow start 0.2 seconds from power-on
- Output voltage regulation: +/- 1%
over the range covered by the primary
at constant load.
- Maximum module/lighting distance
30m with wire of cross-section 6mm
2
Electrical (controls)
-
Potentiometer/switch : 10Kohms, travel
W
300°
-
Diode power supply voltage: 2 VDC
-
Switch to batteries indicator (if WPS 24
option fitted), by red diode
-
Maximum control/module distance:
50m with wire of cross-section 0.6mm
2
2.2.2. Details of WPS 24 backup card
Computation time <0.5s
Dimension 130 x 55
Protection by F20A automobile type fuse
Max power: 400W
Supply voltage 26VDC ± 4VDC
insulation:
V
to 28
AC+DC
rms
V
AC+DC
rms
Ref. 060123001
ENERGIX WPS
2.2.Características
2.2.1. Detalle de los módulos (Figura 2)
WPS 362 / WPS 361 - Versión A
y B
Mecánicas
Térmicas
Eléctricas
Eléctricas
V
AC+DC
rms
V
Eléctricas
2.2.2. Detalle de la tarjeta emergencia
WPS 24
V
AC+DC
rms
AC+DC
rms
2
W
2
9