3. Declaration of Conformity
According to Machinery Directive (2006 / 42 / EC) attachment II 1.A
Herewith we declare that the machine described below is in its conception and design and in
the shape delivered by us in accordance with the fundamental requirement for safety and health
as prescribed in the applicable EC-guidelines. In the case of any change or modification of the
machine not authorised by us this declaration becomes invalid.
Designation of machine:
Drufosmart
D3200 / D3200A / D32001 / D32001A
Type of machine:
Pressure thermoforming unit
Applicable EC directives:
2006 / 42 / EC
Machinery Directive
2014 / 35 / EU
Low Voltage Directive
2014 / 30 / EU
EMV Directive
Applied harmonised standards:
2006 / 42 / EC
According to Machinery Directive
EN ISO 12100:2010
Safety of machinery – General principles for design – Risk assess-
+ Corrigendum to
ment and risk reduction (ISO 12100:2010); German version EN ISO
DIN EN ISO 12100:
12100:2010 + Corrigendum to DIN EN ISO 12100:2011-03.
2011-03
EN ISO 13732-1:2008
Ergonomics of the thermal environment – Methods for the assessment
of human responses to contact with surfaces – Part 1: Hot surfaces
(ISO 13732-1:2006); German version EN ISO 13732-1:2008-12.
DIN EN 60204-1:
Safety of machinery – Electrical equipment of machines – Part 1:
2006 / A1:2009
General requirements (IEC 60204-1:2005/A1:2008); German version
+ Corrigendum to
EN 60204-1:2006/A1:2009 + Corrigendum to EN 60204-1:2006;
DIN EN 60204-1:2006
German version CENELEC-Cor.: 2010 to EN 60204-1:2006.
2014 / 35 / EU
Low Voltage Directive
EN 61010-1:2010
Safety requirements for electrical for measurement, control and Iabo-
ratory use – Part 1: General requirements (IEC 61010-1:2010 + Cor.:
2011); German version EN 61010-1:2010.
2014 / 30 / EU
EMV Directive
EN 61000-3-2:2014
Electromagnetic compatibility (EMC) – Part 3-2: Limits – Limits for
harmonic current emissions (equipment input current <= 16 A per
phase) (IEC 61000-3-2 :2014); German version EN 61000-3-2:2014.
EN 61000-3-3:2013
Electromagnetic compatibility (EMC) – Part 3-3: Limits – Limitation
of voItage changes, voItage fluctuations and flicker in public Iow-
voltage supply systems, for equipment with rated current <= 16 A per
phase and not subject to conditinoal connection (IEC 61000-3-3:2013;
German version EN 61000-3-3:2013.
EN 61326-1:2013
Electrical equipment for measurement, control and laboratory use –
EMC requirements; Part 1: General requirements (IEC 61326-1: 2012);
German version EN 61326-1:2013. Emission according to living area,
busness and industrial undertakings as well as small enterprises; inter-
ference stability according to industrial area.
Dreve Dentamid GmbH
Max-Planck-Straße 31 · 59423 Unna/Germany · www.dentamid.dreve.de
24
4. Safety advice
Attention! Read the following notes before connecting and putting the unit into
service. The operating service and the function of the unit can only be guaranteed for,
if the general safety and accident prevention instructions of the legislation as well as
the safety instructions of the working instructions are observed.
1. This unit must only be used accord-
and understood the manual. Only in
ing to the present working instructions.
that case the operators may operate
We cannot be held liable for damages
the unit.
which have been caused by improper
treatment resp. faulty manners of oper-
8. The unit has to be examined before
ation.
operation in view of proper condition
and operating safety. If the unit is not
2. Put the unit on a stable and even
in proper condition, it must not be used
basis which is suitable for the operation
and has to be marked accordingly.
weight of approx. 16 kg of the unit.
9. The pressure of the compressed air
3. In order to avoid the intrusion of
line must not exceed 10 bars.
water into the unit (e. g. splash water)
10. Do not intrude any objects into
the unit is to be placed into a dry sur-
rounding.
the unit.
4. Do not store any flammable sub-
11. Do not lock-up the operation ele-
stances in the close surrounding of
ments of the unit.
the unit.
12. Do not block the downwards-
5. The indicated voltage on the type
moving piston.
plate has to meet the voltage of the
13. Do not touch the downwards-mov-
power source.
ing piston.
6. The unit must only be connected
with a socket with protective conductor.
14. The swivelling housing of the radia-
Never touch the plug with wet hands.
tor reaches high temperatures and must
therefore not be touched – not even for
7. Authorised operators: The operating
short moments.
company of the machine has to hold
the working instructions at the oper-
15. If the unit is not used, please dis-
ators' disposal and has to make sure
connect the mains.
that authorised operators have read