Aqualytic AL100 Mode D'emploi page 32

Masquer les pouces Voir aussi pour AL100:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 36
Calibration Mode
GB
RESULT
CAL
Mode
Zero
Test
CAL
RESULT + x
On
Off
:
:
32
The result is shown in the display, alternating with CAL.
If the reading corresponds with the value of the calibration standard
(within the specified tolerance), exit calibration mode by pressing the
[ON/OFF] key.
Changing the displayed value:
Pressing the [MODE] key once increases the displayed value by 1 digit.
Pressing the [ZERO/TEST] key once decreases the displayed value by
1 digit.
Press the corresponding key until the reading equals the value of the
calibration standard.
By pressing the [ON/OFF] key, the new correction factor is calculated
and stored in the user calibration software.
Confirmation of calibration (3 seconds).
AL100_6a 05/2019

Publicité

Table des Matières
loading

Table des Matières