Methods

GB
AL
Tablet
Powder
0.0.0
Tablet
Powder
Zero
!
Test
AL
Tablet
Powder
RESULT
Tablet
Powder
26

Aluminium with Tablet

0.01 – 0.3 mg/l
Fill a clean vial (24 mm Ø) with 10 ml of the water sample and
perform zero calibration (see "Operation").
Add one ALUMINIUM No. 1 tablet straight from the foil to the 10 ml
water sample and crush the tablet using a clean stirring rod (dissolve
the tablet).
Add one ALUMINIUM No. 2 tablet straight from the foil to the same
water sample and crush the tablet using a clean stirring rod.
Close the vial tightly with the cap and swirl gently several times until
the tablet is dissolved.
Place the vial in the sample chamber making sure that the
are aligned.
Wait for a reaction period of 5 minutes.
(Countdown can be activated, see page 25)
The method symbol flashes for approx. 3 seconds.
The result is shown in the display in mg/l Aluminium.
Tolerance: ± 0.05 mg/l Al
marks