Menu options

GB
Mode
On
Off
!
Store
Date
Print
Time
2
1
Cal
Store
Date
Print
Time
diS
1
Mode
Zero
Test
Mode
!
Store
Date
Print
Time
Prt
1
PrtG
On
Off
30

Menu selections

Press the [MODE] key and hold.
Switch the unit on using the [ON/OFF] key.
Allow the 3 decimal points to be displayed before releasing the
[MODE] key.
The [!] key allows for selection of the following menu points:
diS recall stored data
1
Prt printing stored data
1
setting the date and time
2 3
Cal
user calibration
diS – Recall of stored data
1
After confirming the selection with the [MODE] key the photometer
shows the last 16 data sets in the following format (automatically
proceeds every 3 seconds until result is displayed):
Number
n xx (xx: 16...1)
Year
YYYY (e.g. 2014)
Date
mm.dd (monthmonth:dayday)
Time
hh:mm (hourhour:minuteminute)
Test
Method
Result
x,xx
The [ZERO/TEST] key repeats the current data set.
The [MODE] key scrolls through all stored data sets.
Quit the menu by pressing [!] key.
Prt – Transmitting stored data (to Printer or PC)
1
Note: To print data, or to transmit to a PC, the optional IRiM (Infrared
Interface Module) is required.
The IRiM Module and the connected printer/PC must be ready. Press
the [MODE] key to start the transmitting, the instrument displays
"PrtG" (Printing) for approx. 1 second followed by the number of the
first data set and its transmission. All data sets will be transmitted one
after the other. After finishing the instrument switches to test mode.
The print job can be cancelled by pressing the [On/Off] key. The
instrument switches off.
If the instrument is not able to communicate with the IRiM, a timeout
AL100_6a 05/2019