Select the future location based on the following criteria:
 No high humidity (see technical data).
 Normal room temperature (see technical data).
 Keep air vents free.
 Not suitable for use with flammable mixtures (such as anaesthetics, oxygen or disinfectants....).
Remove the device from the packaging and place it onto a sufficiently large and horizontal surface.
Do not yet connect the power plug.
Before initial operation, insert the halogen lamp. The device cover must be removed for this
purpose.
 Press the release button (2).
 Remove the cover (1).
During the following operating steps, do not touch the quartz envelope of the inside of the
reflector! This could lead to a premature failure of the lamp.
 Remove all transport packaging (e.g. carton for lamp socket).
 Fit lamp socket (9) on contact pins of the lamp without tilting.
 Insert lamp (8) into lamp holder (10) without tilting and push to the stop. If the lamp features an align-
ment feather, observe the correct engagement.
 Attach the cover (1) (self-engaging).
 Insert the fibre-optic light guide in the connection (4) to the stop. For fibre-optic light guides with feather
pin, it must be inserted in the key to the stop. Ensure that the rotary knob (5) is in position "Off"
(0-position).
 Insert the power plug.
 Activate the device by turning the rotary knob (5) clockwise.