Electrical Generation, Sensor Measurement and Simulation
Frequency Generation 0.0005 ... 10000 Hz
Pulse Generation
Uncertainty includes reference standard uncertainty, hysteresis, non-linearity, repeatability and typical long term stability for men-
1)
tioned period. (k=2)
Range
Resolution
0.0005 ... 0.5 Hz
0.000001 Hz
0.5 ... 5 Hz
0.00001 Hz
5 ... 50 Hz
0.0001 Hz
50 ... 500 Hz
0.001 Hz
500 ... 5000 Hz
0.01 Hz
5000 ... 10000 Hz
0.1 Hz
Feature
Temperature coeffi cient
Maximum load current
Output amplitude positive
square wave
Output amplitude sym-
metric square wave
Duty Cycle
Jitter
Supported units
Feature
Range
Resolution
Maximum load current
Output amplitude
positive pulse
Output amplitude
symmetric pulse
Pulse frequency
Duty Cycle
1 Year Uncertainty
0.01% RDG
0.01% RDG
0.01% RDG
0.01% RDG
0.01% RDG
0.01% RDG
Specifi cation
Specifi cation valid from –10 to 50°C (14...122°F)
5 mA
0 .. 12 Vpp ±(0.2 V+5%)
0 .. 6 Vpp ±(0.2 V+5%)
1 ... 99 % (0.0009 ... 500 Hz),
high/low time: minimum 25μs, maximum 1165 s
< 0.28 μs
Hz, kHz, cph, cpm, 1/Hz (s), 1/kHz (ms), 1/MHz (μs)
Specifi cation
0 to 9 999 999 pulses
1 pulse
5 mA
0 .. 12 Vpp ±(0.2 V+5%)
0 .. 6 Vpp ±(0.2 V+5%)
0.0005 ... 10000 Hz
1 ... 99 % (0.0009 ... 500 Hz),
high/low time: minimum 25 μs, maximum 1165 s
(1
(±)
133