Electrical Generation, Sensor Measurement and Simulation
Electrical Generation, Sensor Measure-
ment and Simulation
mV Measurement (T/C-Terminals) –25 ... 150 mV
mV Generation (T/C-Terminals) –25 ... 150 mV
Voltage Generation –3 ... 12 V
Uncertainty includes reference standard uncertainty, hysteresis, non-linearity, repeatability and typical long term stability for men-
1)
tioned period. (k=2)
The following features are available in model MC2-MF and model MC2-TE.
Range
Resolution
-25 ... 150 mV
0.001mV
Feature
Temperature coeffi cient
Input impedance
Supported units
Display update rate
Range
Resolution
-25 ... 150 mV
0.001mV
Feature
Temperature coeffi cient
Maximum load current
Load effect
Supported units
Range
Resolution
-3 ... -0.25 V
0.1 mV
± 0.25 V
0.01 mV
0.25 ... 12 V
0.1 mV
Feature
Temperature coeffi cient
Maximum load current
Load effect
Supported units
1 Year Uncertainty
0.02 % RDG + 4 μV
Specifi cation
Add 0.0015 % RDG /°C to spec. outside 18 ... 28°C
Add 0.0008 % RDG /°F to spec. outside 64.4 ... 82.4°F
> 10 Mohm
V, mV, μV
3 / second
1 Year Uncertainty
0.02 % RDG + 4 μV
Specifi cation
Add 0.0015 % RDG /°C to spec. outside 18 ... 28°C
Add 0.0008 % RDG /°F to spec. outside 64.4 ... 82.4°F
5 mA
< 5μV/mA
V, mV, μV
1 Year Uncertainty
0.02 % RDG + 0.1 mV
0.02 % RDG + 0.1 mV
0.02 % RDG + 0.1 mV
Specifi cation
Add 0.0015 % RDG /°C to spec. outside 18 ... 28°C
Add 0.0008 % RDG /°F to spec. outside 64.4 ... 82.4°F
5 mA
< 50 μV/mA
V, mV, μV
(1
(±)
(1
(±)
(1
(±)
131