Thermocouple Types Available As Option; Internal Reference Junction - BEAMEX MC2 Série Guide D'utilisation

Table des Matières

Publicité

Thermocouple Types Available as Option

K+/Au0.07Fe
Ni18Mo/Ni0.8Co
Platinel II

Internal Reference Junction

Uncertainty includes reference standard uncertainty, hysteresis, nonlinearity, repeatability and typical long term stability for mentioned
1)
period. (k=2). Uncertainty does not include reference junction uncertainty.
IEC 584, NIST MN 175, BS 4937, ANSI MC96.1
2)
±0.02 % of thermovoltage + 4 μV
3)
DIN 43710
4)
ASTM E 988 - 96
5)
ASTM E 1751 - 95e1
6)
GOCT 3044-84
7)
Temperature Measurement and Simulation
Type T
(2
, -270 ... 400 °C
Range
Resolution
-270 ... -200 °C
0.01 °C
-200 ... 100 °C
0.01 °C
100 ... 400 °C
0.01 °C
Type U
(4
, -200 ... 600 °C
Range
Resolution
-200 ... -100 °C
0.01 °C
-100 ... 600 °C
0.01 °C
(4
Type L
, -200 ... 900 °C
Range
Resolution
-200 ... 900 °C
0.01 °C
(5
Type C
, 0 ... 2315 °C
Range
Resolution
0 ... 1000 °C
0.01 °C
1000 ... 2000 °C
0.01 °C
2000 ... 2315 °C
0.01 °C
(6
Type G
, 0 ... 2315 °C
Range
Resolution
0 ... 100 °C
0.01 °C
100 ... 2315 °C
0.01 °C
(5
Type D
, 0 ... 2315 °C
Range
Resolution
0 ... 1000 °C
0.01 °C
1000 ... 2000 °C
0.01 °C
2000 ... 2315 °C
0.01 °C
(6
(6
(6
Range
1 Year Uncertainty
-10 ... 50 °C
0.25 °C
1 Year Uncertainty
(3
0.3 °C
0.2 °C
1 Year Uncertainty
0.3 °C
0.2 °C
1 Year Uncertainty
0.25 °C
1 Year Uncertainty
0.4 °C
0.8 °C
1.2 °C
1 Year Uncertainty
(3
1.0 °C
1 Year Uncertainty
0.4 °C
0.8 °C
1.2 °C
Au/Pt
(6
Pt40Rh/Pt20Rh
(6
BP (A)-1
(7
(1
(±)
All specifi cations are subject to change without prior notice
(1
(±)
(1
(±)
(1
(±)
(1
(±)
(1
(±)
(1
(±)
Ir40Rh/Ir
(6
Pt5Mo/Pt0.1Mo
(6
XK (L)
(7
137

Publicité

Chapitres

Table des Matières
loading