O P E R A T I O N
GB
MAINTENANCE OF THE CHUCK
Remove the Chuck:
4
Remove chuck out of the shaft, check it, clean it or replace it.
Damaged and fooled chucks reduce the tension and the exact truth of
rotation.
Risk of Accident!
If you work with a defective chuck a risk of accident could be possible.
Cleaning of the shaft:
5
Remove dirt with the cleaning brush MU 0492.
Cleaning the Chuck:
6
Clean the chuck MU 0221 (Ø 2,35 mm) with clean brush MU 0492.
Ø 3,00 mm chucks have the order number MU 0221 B.
Insert the Chuck:
7
Lock shaft of the motor with the box spanner MU 4136 and screw in
the chuck with the chuck key MU 0222 (direction of arrow).
Screw on cap MU 4198.
Mounting of Motor End Cap:
8
Carefully place pins into the contacts.
4164 and tighten it by hand (direction of arrow).
9
Tightening of the Bur:
Insert the bur and turn handpiece housing (direction of arrow).
Just use unobjectionable burs.
Please take a look at the warranty conditions which will be enclosed separately.