10.2 Errors

In case of error alarm the device has to be reset to start a new cycle.
Display
22
 
ALARM RESET
Causes
Water infeed anomaly
- cock closed
- filters obstructed
- water system pressure
anomaly
Water level anomaly
- the device has been
moved or tilted during the
cleaning.
- many tools have been
added after water infeed.
Water drain anomaly
- filters obstructed.
- drain hose in a wrong
position.
- Wrong installation: anti-
return valve on the drain.
Hold down for 8s. buttom OK
Solutions
- Make sure the cock is open.
- Clean tank filter, drain filter, and water
infeed filter.
- Make sure water pressure ranges
within
indicated values. In case of
insufficient
pressure
plumber.
-
Make
sure
the
(OPTIONAL) is not locked.
- Make sure there is no water into the
tank and start a new cycle
- Make sure there is no water into the
tank and start a new cycle.
- Do not add tools during the cycle.
- Clean the tank filter and the drain
filter.
- Position the drain hose correctly, see
installation.
- Make sure there is no water into the
tank and start a new cycle.