6.1

Connections

Take the transparent drain hose to a visible point at the same height of the appliance
leaning plane or under it. Do not obstruct nor keep it immerged into the fluid. The drain
hose has also an overflow function, when the detergent reservoir is full, if some more
detergent is injected the drain hose discharges it.
Check the integrity of the power cord.
 
Connect the power cord to the socket on the rear side of the appliance (M).
Connect the hose to the appliance (O) fitting also
the rubber gasket. Connect the other side to the
feeding cock.
OPTIONAL: emergency water block system
(WATER BLOCK). Such device shall be vertically
mounted immediately downstream the feeding
cock. See the device instructions for further
information on the tool installation and use.
Remove the safety cap. Connect the hose to the
device (P). Connect the other end to the
discharge.
consequently the drain hose has to be lower than
the device. The hose has not to be bent upward
(siphons). Non-return valves must not be present.
Detergent reservoir drain hose
Power cord
Remove safety caps.
Water feeding hose.
Water drain hose.
The
device
discharges
ENGLISH