en - Areas of application

Handles

Irrigation sleeves

Shafts and tool
inserts
Shafted instru‐
ments which can‐
not be dismantled
Trocars with
lumens
40
Fit contaminated handles onto an injector nozzle, e.g. E 452.
E 442 irrigation sleeves are suitable for reprocessing hollow instru‐
ments and instruments with an external diameter less than 10 mm.
Place the shafts of MIS instruments in irrigation sleeves with caps.
Place HF forceps and electrodes in irrigation sleeves with caps,
with the functional ends pointing downwards.
Place instruments which cannot be dismantled in an irrigation slee‐
ve. The functional ends must be open; if necessary use an E 472
spring clamp to hold them open.
Attach the side connectors for instruments with an
E 448 hose adapter. Secure the hose adapter to the injector mani‐
fold using a silicone holder or with a screw fitting.
Push the trocar with lumen into the irrigation sleeve.
Push trocar sleeves of less than 10 mm Ø into irrigation sleeves. At‐
tach the side connector using an E 448 hose adapter. Secure the
hose adapter to the injector manifold using a silicone holder or with
a screw fitting.