Trouble Shooting Information
Should you encounter problems operating your spirometer, please
consult the table below:
Problem
Micro I cannot be switched
on
Every time you switch the
instrument on the time is
shown as 00:00
Micro I is outside of % when
conducting the calibration
check
Electromagnetic Compatibility (EMC)
Changes or modifications to the Micro I spirometer that are not
expressly approved by Carefusion can cause EMC issues with this or
other equipment. The device is designed and tested to comply with
applicable regulation regarding EMC and needs to be installed and
put into service as per interference exceeding levels specified in EN
50082-1:1992
WARNING: use of portable phones or other radio frequency (RF)
emitting equipment near the system may cause unexpected
adverse operation
Possible
Cause
Batteries are
flat
The internal
battery of Micro
I is defective
There are leaks
in the syringe
or connections
32
Solution
Recharge batteries
Contact your
dealer.
Check the syringe
and connections for
leaks