8.1
Tableaux 1:
Immunity (all devices)
Guidance and manufacturer's declaration – electromagnetic immunity
The handy vaq is intended for use in the electromagnetic environment specified below.
The customer or the user of handy vaq hould assure that it is used in such an environment.
Immunity test
standard
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient / burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short interruptions and
voltage variations on power supply lines
IEC 61000-4-11
Note: U
is the a.c. mains voltage prior to application of the test level.
T
Power frequency (50 / 60 Hz)
magnetic field
IEC 61000-4-8
IEC 60601
Compliance
test level
level
± 8 kV contact
± 8 kV contact
±15 kV air
±15 kV air
± 2 kV for power sup-
ply lines
± 2 kV for power sup-
ply lines
± 1 kV for
input/output lines
± 1 kV line to line
± 1 kV line to line
± 2 kV line to earth
± 2 kV line to earth
<5% U
(0,5 cycle)
<5% U
T
T
40% U
(5 cycles)
40% U
T
T
70% U
(25 cycles)
70% U
T
T
<5% U
for 5 s
<5% U
T
T
3 A / m
200 A / m
Mode d'emploi
handy vaq
Electromagnetic
environment - guidance
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%
Mains power quality should be that of a ty-
pical commercial or hospital environment.
Mains power quality should be that of a ty-
pical commercial or hospital environment.
(0,5 cycle)
(5 cycles)
Mains power quality should be that of a ty-
pical commercial or hospital environment.
(25 cycles)
for 5 s
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or hospital
environment.
page 35