2. Press the start button.
Note:
• Your unit has been designed to automatically output
the highest level or airflowThe airflow.
• The Airflow Volume Indicator Light will illuminate at
the highest setting.
To adjust the airflow, and nebulizer rate
1. Your Ulrasonic Nebulizer has 3
airflow setting: Maximal, Medium
and Minimum.
2. Based on instruction from your
physician and/or respiratory the-
rapist, press the Nebulization rate
control button to adjust the air-
flow volume. The Airflow Volume
Indicator Light will show the level
(Maximal, Medium or Minimum)
Fig. 5
selected, Fig. 5.
3. The Transfer Cone located inside
the Transducer enhances the level is too strong, the cone
should be carefully removed after the unit has comple-
tely cooled. Restart the unit and resume treatment.
To begin a treatment
1. Follow the instruction provided by your physician and/
or respiratory therapist.
2. Your Nebulizer is designed to start at a lower atomi-
zation rate in the beginning and will become stronger
after a few minutes.
Remember:
• Keep your eyes away from the output of medication
mist.
• If any discomfort or abnormality occurs, stop using
the unit immediately.
• Do not tilt or shake the unit when in operation ..
• Opening the top cover during operation may cause a
severe injury.
To end a treatment
1. When the treatment has ended, press the power but-
ton to shut off the unit. Otherwise,your unit will auto-
matically shut off after approximately 10 minutes of
continuos use.
2. Unplug the adapter from the outlet and detach the
adapter plug from unit's AC port.
3. Allow the unit to cool - approximately 20-30 minutes.
Remember:
• The unit should not be used for more than 20 minu-
tes at the time.
• Wait 30 minutes before reusing the unit.
After every use
1. Allow the unit to completely cool.
2. Remove the top cover.
3. Carefully remove the medicine cup and pour remai-
ning medication.
4. Empty the water tank.
5. Wipe out the water tank and transducer with a soft,
lintfree cloth.
6. Follow the cleaning procedures provided in this guide-
book.
ENGLISH
TROUBLESHOOTING
If any abnormality occurs during use, please check
and correct the following:
Unit does not operate when power button is pressed:
• Check the Ac adapter connection to the outlet and/or
check the AC plug to the unit's AC port.
• Check that the water tank is filled to the correct level.
• If using with optional battery pack, check that the bat-
tery is completely charged. Check that the battery is
mounted correctly to the unit.
No misting or low rate of misting
• Check that there is medication in the medicine cup.
• Check the medication cup for damage.
• Check that the transducer is clean.
Replacement Parts
Replacement parts are available through your local retai-
ler or they can be ordered by mail, See separate replace-
ment part of list for details.
Your PINGOO Ultrasonic Nebulizer can operate with
the use of a rechargeable battery when AC power is not
available.The life of the battery life will vary depending on
usage, storage and charging conditions.
To charge the battery
1. Plug the adapter into standard AC outlet.
2. Insert the adapter plug into the unit's AC port, Fig. 1
Note:
• For first time use or after storing
the battery for a period time, char-
ge the battery at least 24 hours,
and thereafter 4 to 12 hours de-
pending on the length of usage.
Fig. 1
• A fully charged battery will last for 30-40 minutes of
continuos use.
• To prolong the life of the battery, the charging current is
designed at the minimum level. Therefore, it is recom-
mended to charge the battery after each use and let it
charge overnight.
To use the battery
1. Insert the battery's plug into nebu-
lizer's power receptacle. Fig.1.
2. Begin the use the Nebulizer as de-
scribed in the Operating Your Ne-
bulizer section of this guidebook.
3. If the nebulizer's indication lamp
changes to red color, it indicates
that the battery is low voltage
Fig. 1
condition and needs to be char-
ged.
4. After use, dispose excess medi-
cation and water before discon-
necting the battery.
5. Always remove the battery from
nebulizer whenever not in use.