The device is intended for use in the electromagnetic environment specified below.
The customer or the user of the device should assure that it is used in such and environment.
Immunity test
Conducted RF
3 Vrms:
IEC 61000-4-6
0,15 MHz – 80 MHz
6 Vrms:
in ISM and amateur
radio bands between
0,15 MHz and 80 MHz
80 % AM at 1 kHz
Radiated RF
10 V/m
IEC 61000-4-3
80 MHz – 2,7 GHz
80 % AM at 1 kHz
NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur
radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which the device is used exceeds the applicable RF compliance level above, the device should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the device.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or
the user of the device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the device as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum output power
of transmitter
W
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated
using the equation applicable to the frequency of the transmitter, where p is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
NOTE1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
Manufacturer's declaration-electromagnetic immunity
IEC 60601 test level
Not applicable
Not applicable
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
Recommended separation distance between
portable and mobile RF communications equipment and the device
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d =1,2 √P
N/A
N/A
N/A
N/A
N/A
Compliance level
Portable and mobile RF communications
equipment should be used no closer to any part of
the device including cables, than the recommended
separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance:
d = 1,2 √P
d = 1,2 √P 80MHz to 800 MHz
d = 2,3 √P 800MHz to 2,7 GHz
Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in metres (m).
Interference may occur in the vicinity of equipment
marked with the following symbol:
m
150 kHz to 80 MHz
d =1,2 √P
N/A
N/A
N/A
N/A
N/A
EN-16
Electromagnetic environment-guidance
150 kHz to 80 MHz
d =1,2 √P
N/A
N/A
N/A
N/A
N/A