MD 9040 & MD 9050 Digital multimeter
AC & AC+DC Voltage
Function RANGE
50Hz ~ 60Hz
mV
60.00mV,
600.0mV
V
9.999V, 99.99V,
999.9V
40Hz ~ 500Hz
mV
60.00mV,
600.0mV
V
9.999V, 99.99V
999.9V
500Hz ~ 1kHz
mV
60.00mV,
600.0mV
V
9.999V, 99.99V
999.9V
1kHz ~ 3kHz
mV
60.00mV,
600.0mV
V
9.999V, 99.99V,
999.9V
3kHz ~ 20kHz
mV
60.00mV
600.0mV
V
9.999V, 99.99V
999.9V
1)
Specificied from 30% to 100% of range.
CMRR:
>60dB @ DC to 60Hz,
Rs=1k
Input Impedance: 10M, 50pF nominal
(80pF nominal for 600mV range)
Residual reading less than 5 digits with
test leads shorted.
Audible Continuity Tester
Audible threshold: between 20 and
300; Response time < 100s
TM
AutoCheck
(ACV)
1)
RANGE
50Hz ~ 60Hz
9.999V, 99.99V, 999.9V
Lo-Z ACV Threshold:
> 3VAC (50/60Hz) nominal
Lo-Z ACV Input Impedance:
Initially approx. 3.0k, 150pF nominal;
Impedance increases abruptly within a
Accuracy
0.5% + 3d
0.8% + 4d
1.0%+4d
2.0%+4d
2.0% + 3d
1.0%+4d
2.0%+4d
2%+3d
3.0%+4d
1)
,
2%+3d
1)
3dB
Unspec'd
Accuracy
1.0%+4d
fraction of a second as display voltage
is above 50V (typical). Ended up
impedances
typically are:
18k
@100V
125k
@ 300V
320k
@ 600V
460k
@ 1000V
DC Voltage
Function RANGE
mV
60.00mV
600.0mV
V
9.999V, 99.99V,
999.9V
NMRR:
>60dB @ 50/60Hz
CMRR:
>110dB @ DC, 50/60Hz,
Rs=1k
Input Impedance: 10M, 50pF nominal
(80pF nominal for 600mV range)
dBm
At 600, -11.76dBm to 54.25dBm,
Accuracy: 0.25dB + 2d (@40Hz --
20kHz)
Input Impedance: 10M, 50pF nominal
Selectable reference impedance of 4, 8,
16, 32, 50, 75, 93, 110, 125, 135, 150,
200, 250, 300, 500, 600, 800, 900, 1000,
1200
TM
AutoCheck
(DCV)
RANGE
9.999V, 99.99V, 999.9V
Lo-Z DCV Threshold:
> +1.5VDC or < -1.0VDC nominal
Lo-Z DCV Input Impedance:
Initially approx. 3.0k, 165pF nominal;
Impedance increases abruptly within a
fraction of a second as display voltage
is above 50V (typical). Ended up
impedances
typically are:
18k
@100V
125k
@ 300V
320k
@ 600V
500k
@ 1000V
21
Specification
vs
display
voltages
Accuracy
0.12%+2d
0.06%+2d
0.08%+2d
Accuracy
0.5%+3d
vs
display
voltages