MD 9040 & MD 9050 Digital multimeter
6 Specification
General Specifications
Display:
Polarity:
Update Rate:
Low Battery:
Operating Temperature:
Relative Humidity:
Pollution degree:
Storage Temperature:
Altitude:
Temperature Coefficient:
Sensing:
Safety:
Transient protection:
MD 9040 & MD 9050 Terminals (to COM) Measurement Category:
V / A / mAA :
Overload Protections:
MD 9050, MD 9040:
A & mA:
A:
V, mV, , & Others:
E.M.C. :
In an RF field of 3V/m:
Capacitance function is not specified
Other function ranges:
Total Accuracy = Specified Accuracy + 100 digits
9999 counts: ACV, DCV, Hz & nS
6000 counts: mV, A, mA, A, Ohm & Capacitance
Automatic
Digital Display: 5 per second nominal;
41 Segments Bar-graph: 60 per second max
Below approx. 7V
o
o
0
C to 45
C
Maximum relative humidity 80% for temperature up to 31
decreasing linearly to 50% relative humidity at 45
2
o
o
-20
C to 60
C, < 80% R.H. (with battery removed)
Operating below 2000m
nominal 0.15 x (specified accuracy)/
o
o
28
C ~ 45
C), or otherwise specified
MD 9050: AC+DC True RMS
MD 9040: AC True RMS
Double insulation IEC/UL/EN61010-1 Ed. 3.0, IEC/EN61010-
2-030
Ed.
1.0,
IEC/UL/EN61010-031 Ed. 1.1 and CAN/CSA-C22.2 No.
61010-1-12 Ed. 3.0 to Category IV 1000Vac & Vdc.
12kV (1.2/50s surge)
Category IV 1000Vac & Vdc
0.44A/1000Vac & Vdc, IR 10kA or better, F fuse
11A/1000Vac & Vdc, IR 20kA or better, F fuse
1050Vrms, 1450Vpeak
Meets EN61326-1:2006 (EN55022, EN61000-3-2, EN61000-
3-3, EN61000-4-2, EN61000-4-3, EN61000-4-4, , EN61000-
4-5, EN61000-4-6, EN61000-4-8, EN61000-4-11)
19
o
o
C @(0
C ~ 18
IEC/EN61010-2-033
Specification
o
C
o
C
o
C or
Ed.
1.0,