TROUBLESHOOTING

In case of problems with the device, please refer to the following table.
Problem
Possible causes
Extremely low
Medication chamber is not securely
spray rate
installed
Medication does not get on the
membrane for more than 15 seconds
Medication chamber membrane is
clogged
Medication chamber contacts are
contaminated with medication or water
Nebulizer and medication chamber
contacts are contaminated
After switching
Medication chamber is not securely
on, the power
installed
indicator lights
Medication chamber is empty
up for 1 second
and then goes
Medication does not get on the
out immediately
membrane
Nebulizer and medication chamber
contacts are contaminated
Power indicator
Batteries are improperly installed
does not light
and the nebulizer
Low battery level
does not work
USB-C cable is not properly connected
to the nebulizer
Power indicator
If the low battery indicator is on, the bat-
is on and the
teries are empty or their capacity is not
nebulizer does
sufficient to operate the device
not work
Medication chamber membrane is
damaged
Medication chamber contacts are con-
taminated with medication or water
Nebulizer and medication chamber con-
tacts are contaminated
Medication chamber membrane is
heavily clogged
74
Solution
Install the medication chamber properly and start
spraying
Place the nebulizer at an angle so that the
medication gets on the membrane
Clean the medication chamber.
If this does not solve the problem, replace the
chamber with a similar new one
Clean the contacts from medication and water
residues and start spraying again
Clean the contacts and restart the device
Install the medication chamber properly and start
spraying
Fill in the medication chamber
Place the nebulizer at an angle so that the
medication gets on the membrane
Clean the contacts and restart the device
Re-install the batteries, observing the polarity,
and restart the device
Replace the batteries and restart the device
Install properly and restart the device
Replace the batteries and restart the device
Replace the medication chamber with a new one
and fill in the medication
Clean the contacts from medication or water resi-
dues and restart the device
Clean the contacts and restart the device
If it is still not possible to use after cleaning, re-
place the medication chamber with a new one