4. Ordering Information

DESCRIPTION
1. Mechanical design
Housing T24 for rail and wall mounting
2. Rated frequency
1) 50 Hz (60 Hz possible without addi-
tional error; 16 2/3 Hz, additional
error 1.25 · c)
2) 60 Hz (50 Hz possible without addi-
tional error; 16 2/3 Hz, additional
error 1.25 · c)
3) 16 2/3 Hz (not reprogrammable by
user, 50/60 Hz possible, but with
additional error 1.25 · c)
3. Power supply
Nominal range
1) AC
90...110 V
2) AC
99...121 V
3) AC
207...253 V
4) AC
360...440 V
5) AC
450...550 V
6) AC
623...762 V
7) DC/AC 24... 60 V
8) DC/AC 85...230 V
4. Power supply connection
1) External (standard)
2) Internal from voltage input
Line 2: Not available for rated
frequency 16 2/3 Hz and applications
A15/A16/A24
Caution! The power supply voltage must
agree with the input voltage (Table 2)!
5. Test certificate
0) None supplied
1) Supplied
6. Programming
0) Basic
9) According to specification
Line 0: Not available if the power supply
is taken from the voltage input
Line 9: The filled form W 2388 e must
be part of the order. If primary values
of measurands and meters are re-
quired you have to fill the form as
well.

5. Technical data

Inputs
Waveform:
Sinusoidal
Rated frequency:
50, 60 or 16 2/3 Hz
MARKING
400 - 1
1
2
3
H
= 100 V
1
n
H
= 110 V
2
n
H
= 230 V
3
n
H
= 400 V
4
n
H
= 500 V
5
n
H
= 693 V
6
n
CSA approved
7
CSA approved
8
1
2
0
1
0
9
Consumption [VA]
(at external power
supply):
Voltage circuit: U
Current circuit: ≤ I
Continuous thermal ratings of inputs
Current circuit
10 A 400 V
Voltage circuit
480 V single-phase AC system
831 V three-phase system
Short-time thermal rating of inputs
Input
Number
variable
of
inputs
Current circuit
400 V single-phase AC system
693 V three-phase system
100 A
5
250 A
1
Voltage circuit 1 A, 2 A, 5 A
Single-phase
AC system
600 V
H
: 1.5 Ur
10
intern
Three-phase
system
1040 V
H
: 1.5 Ur
10
intern
L
W
ON
ORKS
®
Interface
Standard program ID:
80 00 36 15 03 04 04 01
Network protocol:
L
ON
Transmission medium: Echelon FTT-10A transceiver,
transformer coupled, reverse
polarity protected, twisted 2-wire
cable
Distance:
500 m
2700 m (doubly-terminated bus
Network wiring:
(22 to 16 AWG)
0.65 to 1.3 mm
Transmission speed:
78 kBit/s
Node within a subnet:
127
Subnet:
255
2
/ 400 kΩ
· 0,01 Ω
2
single-phase AC system
693 V
three-phase system
Duration
Interval
of
between two
overload
overloads
3 s
5 min.
1 s
1 hour
10 s
10 s
10 s
10 s
T
®
ALK
(free topology)
topology)