10. Error Messages

Display / Problem
26
Display
Meaning
Measured
Temperature
too high
Measured
temperature
too low
Ambient
temperature
too high
Ambient
temperature
too low
Error function
display
Blank display
Empty battery
indication
Possible cause
and fault
remedy
Displays „H" when
measured
temperature higher
than 100.0 °C or
199.9 °F.
Displays „L" when
measured
temperature lower
than 0 °C or 32.0
°F.
Displays „H" in
combination with
the „ " when am-
bient temperature
is higher than 40.0
°C or 104.0 °F.
Display „L" in
combination with
the „ " when
ambient tempera-
ture is lower than
5 °C or 41 °F.
When item has
malfunction. If mal-
function occurs fre-
quently, please con-
tact manufacturer
(see page ??)
Please check if the
battery has been
inserted corecctly.
Also check poliari-
ty (<+> and <–>) of
battery.
If the battery icon
is shown on the
display, the batter-
ies should be
replaced immedi-
ately.