29
Guidance and manufacture's declaration-electromagnetic immunity
The CMS8000 Patient Monitor is intended for use in the electromagnetic environment specified be-
low. The customer or the user of CMS8000 Patient Monitor should assure that it is used in such an
environment
Immunity test
IEC 60601
test level
Conducted RF
3Vrms
IEC61000-4-6
150KHz to 80MHz
Radiated RF
3V/m
IEC61000-4-3
80MHz to 2.5GHz
NOTE 1 At 80MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telepho-
nes and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot
be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in
the location in which the device is used exceeds the applicable RF compliance level above, the de-
vice should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as re-orienting or relocating the device.
b Over the frequency range 150 KHz to 80 MHz, field strengths should be less than 3V/m.
Compliance
Electromagnetic environment -
level
guidance
3Vrms
Portable and mobile RF communica-
tions equipment should be used no clo-
ser to any part of the device, including
cables, than the recommended separa-
tion distance calculated from the equa-
tion applicable to the frequency
3Vrms
of the transmitter.
Recommended separation distance
d= 1.2 √P
d= 1.2 √P 80MHz to 800MHz
d= 2.3 √P 800MHz to 2.5GHz
Where P is the maximum output power
rating of the transmitter in watts (W) ac-
cording to the transmitter manufacturer
and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmit-
ters, as determined by an electroma-
gnetic site survey,
the compliance level in each frequency
range.
Interference may occur in the vicinity of
equipment marked with
the following symbol:
should be less than
a
b