List Of Possible Anomalies And Relative Solutions - Seko Pool Basic Pro Evo Manuel D'installation

Ph-/rx/ph+ - ph/rx/flocc
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

13. LIST OF POSSIBLE ANOMALIES AND RELATIVE SOLUTIONS

ANOMALY
The
instrument
always indicates pH
7.00
The
instrument
always indicates a
high or continuously
unstable value
Impossible
to
calibrate the pH 7
value
Error shown on the
display
Calibration quality of
pH probe < 20%
Impossible
to
calibrate the pH 4
value
Error shown on the
display
Calibration quality of
pH probe < 20%
Slow
electrode
response
000013791
CAUSE
Problem with the cable and/or
connector.
The electrode's connection
cable is damaged.
There is an air bubble in the
electrode's membrane.
Electrode worn.
Connection cable too long or
too close to an electrical wire:
Disturbance.
Unsuitable buffer solution.
Problem on probe's porous
material, dirt deposits.
Electrode worn.
Unsuitable buffer solution.
Problem with the electrode
bulb.
Electrode worn.
Electrode
electrostatically
charged.
Rev. 1.0
SOLUTION
1) Check for possible short circuits on the
electrode  instrument connection cable
(between the cable's core and the external
shielding).
2) Make sure that there are no traces of
humidity
and/or
condensation
connector of the probe or the device.
3) Make sure that 100  resistance is
present between terminals 11 and 12.
Check the cable.
Place the electrode vertically and shake it
slightly until the air bubble rises.
N.B.: The electrode must be vertical or
tilted by a maximum of 45°.
Replace the electrode.
Reduce the distance between the device
and the probe.
Make sure that the solution used is pH 7.
Check the buffer solution pH using an
electronic pH-meter.
Use a new pH 7 buffer solution and restart
calibration.
Make sure that the probe's porous material
is in good condition; wash the electrode
using a diluted acid-based solution and dry
with a soft cloth.
Replace the electrode.
Make sure that the solution used is pH 4.
Check the buffer solution pH using an
electronic pH-meter.
Use a new pH 4 buffer solution and restart
calibration.
Make sure that the electrode bulb is not
damaged. Make sure that it did not become
dry outside of the water. As a last resort,
clean the electrode and leave it immersed
in the water for a few hours.
Replace the electrode.
During the calibration phase, the electrode
MUST NOT be dried with a cloth or paper;
let it drip.