AlcoQuant 6020
7
Error messages and removal of errors
Each device error is signalised acoustically by a short dual signal and optically either by the
yellow LED (error that does not impair measurement) or by the red LED (error impairing
measurement). It leads to a corresponding error message on the display.
The error message must be acknowledged by pressing the "OK" key
Error indication
User error
Blow failure
Device errors which do not endanger the measuring process (yellow LED)
Error buzzer
Error LED red
Error date/time
32
Possible cause
Blow failure
This error appears if during
the blowing-in process the
respiratory flow drops below
the required level before the
blowing volume of at least
1.2 L was reached.
Signal generator
A defective signal generator
has been detected.
Error red LED (green,
yellow)
A defective, or only weakly-
lit LED was detected.
Time
An error was detected while
the current date/time was
being determined.
briefly.
Remedy
After the failure is output (approx. 2
seconds), the device switches back to the
previous measuring mode.
A signal tone and the flashing green LED
blinking at an interval of 1.5 seconds
signalise that the device is ready for
measuring.
The blowing process can be restarted.
The test person should blow longer and
harder than before.
Measuring is still possible
The error message is only output during
the self test (while the device is being
switched on)
If the error recurs, please contact
Service.
Measuring is still possible
The error message is only output during
the self test (while the device is being
switched on)
If the error recurs, please contact
Service.
Measuring is still possible
The correct date/time may not be
displayed. If this is the case, incorrect
date/time values will be stored to memory
with the current measured value.
If the error recurs, please contact
Service.