AlcoQuant 6020
5
Maintenance / Cleaning
Calibration
The maintenance is required at least every 6 months. This includes the following measures:
•
Calibration of the device
•
Test whether device functions are in proper working order
•
Test of calibration
The maintenance must be carried out only by authorised specialists or EnviteC
Service.
Cleaning
Please use only an impregnated disposable cloth with mild cleaning agents to clean the device
without excessive pressure
Disinfection
Please use only an impregnated disposable cloth with Isopropyl-Alcohol to clean the device
without excessive pressure (EN 60601-1) (Check of the permanency of the inscription).
Keep battery cover closed while cleaning and disinfection.
Humidity must not enter the holes of the device.
Care must be taken that the electrical contacts inside the device (battery) and on the
serial interface remain clean and dry to ensure that the device functions perfectly.
If alcohol has been used for cleaning/disinfection, a waiting time of min. 15 minutes
until the next measurement must be maintained!
30
Dok.-Nr.:
059-07-11000300-k.doc
/ 07.08