2) Fill Syringe
a) Remove protective cap from the
primary vial packaging and prepare
the top following aseptic techniques.
b) Aseptically remove the Filling/Vial
Adapter from its packaging.
c) Aseptically remove the Syringe from
its packaging. Remove the blue
protective Syringe Cap.
d) Insert the Syringe into the top of the
Filling/Vial Adapter. Push down and
twist clockwise to lock Syringe into
place, as shown in figure 4.
e) If needed, remove the clear
protective sleeve from the Filling
Adapter, as shown in figure 5.
f) Insert the Filling/Vial Adapter spike
into the vial with a twisting motion
until it is fully inserted, as shown in
figure 6.
g) Invert the assembly and pull down
on the Syringe Plunger to draw
fluid into the Syringe, as shown in
figure 7. Remove air bubbles as
required.
h) Complete filling by aligning the red
rubber ring with the Fill Line, as
shown in figure 8.
i) Snap off the end of the Plunger
protruding from the Syringe, as
shown in figure 9.
NOTE: The Filling/Vial Adapter is
for single vial use. It may be used
to continuously fill several Syringes
from the same multi-dose vial for up
to 48 hours. Up to five vial punctures
with a new Filling/Vial Adapter can be
performed on a multi-dose vial. Cap
the Filling/Vial Adapter with a new un-
used Syringe between fillings.
NOTE: Prior to vial storage, remove
and dispose of the Filling/Vial Adapter
in accordance with local requirements.
figure 4
figure 5
figure 6
figure 7
figure 8
figure 9