InTrac
®
777-SL / InTrac
4.
Installation/Startup and shutdown of manually
and pneumatically operated housings
4.1
Preparation of the equipment
The housings are mounted and fixed on the vessel (reactor, tank,
pipe, etc.) by means of a ring nut in conjunction with a weld-in
socket, by a flange connection or a 1" MNPT external (male)
thread.
Attachment of the weld-in socket or the flange connection to the
vessel is the responsibility of the customer.
Note: Weld-in sockets with thread G 11/4" (internal ø= 25 mm,
L= 40 mm, straight or at an angle of 15°) are available from
your supplier in a choice of materials.
Important! After welding, the bore of the weld-in socket must be
checked and reamed to size 25-H7 if necessary.
To ensure correct functioning of the retractable housing, please
note the following information on positioning:
– The retractable housing can be installed vertically or at an
© 06/00 Mettler-Toledo GmbH, CH-8606 Greifensee
Printed in Switzerland
®
776-SL
angle. In the latter case, ensure that the angle of the hou-
sing is equal to or greater than 15° above the horizontal
(see following illustration).
Flange
Socket
Direct mounting
1" MNPT
Warning! Installation of the retractable housings at other
than the admissible mounting positions is not allowed, since
in such positions proper functioning of the electrode/sensor
cannot be guaranteed.
Flow-through
InTrac 777-SL-04-CH
52 400 841
13