Technical data
Probe
Measuring principle
Measuring range
Accuracy
Resolution
Minimum curvature
Minimum measuring area
Basic thickness of material
Working temperature
Air humidity
Weight
Dimensions
INTENDED USE OF INSTRUMENT
• Non-destructive measurement of thickness of non-magnetic coatings on magnetic and non-magnetic
metals on the principles of magnetic induction or eddy current principle - with one probe only
• Saving of data and reading out the information stored after transferring to PC via USB port
CARE AND CLEANING
• Handle measuring instruments with care.
• Clean with soft cloth only after any use. If necessary damp cloth with some water.
• If instrument is wet clean and dry it carefully. Pack it up only if it is perfectly dry.
• Transport in original container / case only.
F
(magnetic/ferrous
metals)
Magnetic induction
0 - 1250 μm
0 - 49,21 mils
0 - 850 μm (±3 % + 1 μm)
850 - 1250 μm (±5 %)
0 - 33,46 mils
(±3 % + 0,039 mils)
33,46 - 49,21 mils (±5 %)
0 - 50 μm (0,1 μm)
50 - 850 μm (1 μm)
850 - 1250 μm (0,01 mm)
0-1,968 mils (0,001 mils)
1,968 - 33,46 mils (0,01 mils)
33,46 - 49,21 mils (0,1 mils)
1,5 mm
Ø 7 mm
0,5 mm
0°C - 40°C
(32°F - 104°C)
20 % - 90 %
110 Gr.
113,5 x 54 x 27 mm
N
(non-magnetic/-ferrous
metals)
Eddy current
0 - 1250 μm
0 - 49,21 mils
0 - 850 μm (±3 % + 1,5 μm)
850 - 1250 μm (±5 %)
0 - 33,46 mils
(±3 % + 0,059 mils)
33,46 - 49,21 mils (±5 %)
0 - 50 μm (0,1 μm)
50 - 850 μm (1 μm)
850 - 1250 μm (0,01 mm)
0-1,968 mils (0,001 mils)
1,968 - 33,46 mils (0,01 mils)
33,46 - 49,21 mils (0,1 mils)
3 mm
Ø 5 mm
0,3 mm
0°C - 40°C
(32°F - 104°C)
20 % - 90 %
110 Gr.
113,5 x 54 x 27 mm
EN
C
D
17