15 Instant help

In case of a fault in the program sequence, the balance should be shortly switched
off. The weighing process must then be restarted from the beginning.
Failure:
The displayed weight
does not glow.
The displayed weight is
permanently changing
The weighing result is
obviously incorrect
Should other error messages occur, switch balance off and then on again. If the
error message remains inform manufacturer.
46
Possible cause:
 The balance is not switched on.
 The mains supply connection has been interrupted
(mains cable not plugged in/faulty).
 Check fuse of adapter / glowing green LED next to
fuse
 Power supply interrupted.
 Batteries are inserted incorrectly or empty
 No batteries inserted.
 Draught/air movement
 Table/floor vibrations
 The seat surface/weighing plate is in contact with
foreign bodies or is not correctly positioned.
 Electromagnetic fields / static charging (choose
different location/switch off interfering device if
possible)
 The display of the balance is not at zero.
 Adjustment is no longer correct.
 Great fluctuations in temperature.
 The balance is on an uneven surface.
 Electromagnetic fields / static charging (choose
different location/switch off interfering device if
possible)