10
4
Preparing the cables
For optimum results, proceed as follows:
• Remove the outer sheath of the cable and depending on the cable diameter,
expose the braided screen for about 10 ...15 mm.
• Push the cap nut and the bladed insert with the sealing ring onto the cable.
• Bend the braided screen outward at right angles (90°).
• Crimp the braided screen over towards the outer sheath of the cable, i.e. fold
it a further 180° .
• Slip the intermediate connection piece on as far as the braided screen, quickly
turning it to and fro around the cable axis.
• Push the bladed insert with the sealing ring into the intermediate connection
piece and engage the locking element.
• Firmly tighten the cap nut.
Cap nut
Shielded cable
Fig. 4.1: Screwed clamping gland assembly
HBM
Bladed insert
Sealing ring
Intermediate connection piece
Braided screen
VKIA405
A2448-1.1 en/de/fr