6.3
Specifications
6.3.1
UL-Designated Product
Model
Cat No.
Dimensions
Weight
Supply Voltage
Supply Frequency
Mains Inlet Fuses*
Transformer Fuse*
PCB Fuse*
Insulation
Installation
Pollution
Duty Cycle
Rated Load
Disposable Bag Size
Minimum Recommended
Capacity
Maximum Recommended
Capacity
Maximum Recommended
Sample Temperature
Recommended Operating
Ambient Temperature Range
Recommended Operating
Relative Humidity Range
Adjustable Timer Settings
No Load Paddle Speeds
(@ nominal 20°C)
Paddle Clearance
* Only these or equivalent UL-approved fuses should be used.
Issue: 3.2
®
STOMACHER
400 CIRCULATOR
USER INSTRUCTIONS
®
Stomacher
400 Circulator
0400/001/AJ
350mm wide x 335mm deep x 280mm high (unpacked)
23kg (unpacked)
100-120V
50-60Hz
T2.0A 250V 5x20mm, breaking capacity: 100A @ 250V
T250mA 250V 5x20mm, breaking capacity: 35A @ 250V
F2.5A 250V 5x20mm, breaking capacity: 100A @ 250V
Class 1 (Earthed)
Over Voltage Category II
Pollution Degree 2
Continuous
140W
177mm x 304mm
80ml
400ml
60°C
10-35°C
10-89%
0 seconds - 99 minutes 59 seconds (±5%)
75rpm to 300rpm in 5rpm increments (±5%)
0-l0mm
Page 19 of 33