Technical Data
In order to regulate the requirements for EMC with the aim to prevent unsafe product
situations, the EN60601-1-2 standard has been implemented. The nebulizer MB0500400
and MB0500500 conforms to the EN60601-1-2:2007 standard for both immunity and
emissions.
Nevertheless, do not use the nebulizer close to the strong electrical or electromagnetic
fields. This may result in incorrect operation and create a potentially unsafe situation.
Guidance and manufacturer's declaration - of MB0500400 and MB0500500
Guidance and manufacturer's declaration-electromagnetic emissions
The MB0500400 and MB0500500 are intended for use in the electromagnetic
environment specified below.
The customer or the user of the MB0500400 and MB0500500 should assure that it is
used in such an environment.
Emission test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations
/flicker emissions
IEC 61000-3-3
Compliance
Electromagnetic environment-guidance
Group 1
The MB0500400 and MB0500500 uses RF
energy only for its internal function. Therefore,
its RF emissions are very low and are not likely
to cause any interference in nearby electronic
equipment.
Class B
The MB0500400 and MB0500500 is suitable
for use in all establishments, including
domestic establishments and those directly
connected to the public low-voltage power
Class A
supply network that supplies buildings used for
domestic purposes.
Compliance
17