4.3. Check of flow
•
Switch the KITE on by pressing the ON/OFF
button. After +/- 15 seconds a signal sounds
to confirm the correct function of the
electronic control unit.
•
Wait for another 1 minute in order to be sure
that the motor speed is constant.
•
Place the flow meter on the air outlet in such
way that the flow meter is in vertical position.
•
The indicator ball should be above the minimum
level.
If not, the air flow is not sufficient.
•
Warning: placing the flowmeter on the outlet will
increase the flow resistance. The electronic flow
control will start compensating by increasing the
flow. Therefore checking of the flow should not take
more time than 3 seconds.
•
After checking the flow, switch off the KITE.
See section 7 for troubleshooting.
Warning: Do not close the slits of the flow meter by
fingers as this will influence the correct function.
Minimum level
6