5.6 Batteries
Defective batteries can only be replaced by authorised distributor workshops.
5.7 Charger
A defective charger cannot be repaired but has to be replaced by a new one.

5.8 Cleaning and disinfecting

Clean the KITE using a cloth with a mixture of water and a mild detergent. Do not
use any solvents. During cleaning ensure that no water is allowed to enter the
blower unit. The blower unit must not be immersed in fluid.
Use a disinfecting spray like Incidur (Ecolab Healthcare) on the inside of the visor
and face seal to remove bacteria and germs. Let it be on the surfaces for 1 minute.
Dry the KITE with a cloth and/or leave it to dry in the open air.
Warning: do not open the motor unit as all guarantee will be invalidated.
Warning: never modify or alter this product as all guarantee will be invalidated.
5.9 Checking the flow alarm (optional)
Make sure the battery is fully charged.
Place the alarm checking device partnr 30.019
(not included)
clockwise till it locks in the same as a filter.
See 4.2.
Switch the KITE on by pressing the ON/OFF
button. After +/- 15 seconds 4 short beeps
sounds to confirm the correct function of the
electronic control unit.
Wait for another 1 minute in order to be sure that the motor speed is constant.
Close one small hole by using a piece of adhesive tape and wait at least 30 seconds
till the motor speed is constant.
Close a second small hole and wait at least 30 seconds till the motor speed is
constant.
Continue closing the other small holes in the same way till the warning signal is
audible. This is a beep repeated every second.
Now check the flow, see 4.3.
Warning: do not close the big holes.
and secure it by turning
13