(Nebulizer HI-FLO, Adult Mask, Pediatric Mask,
For each individual patient it's recommended to use the nebulizer for 6 months or for a maximum of 120
treatments. The nebulizer must be replaced after a long period of inactivity, if it is deformed or broken, or if
the nebulizer nozzle is blocked by dry medicine, dust, ecc..
Only use the original nebulizer supplied by CA-MI with the device
The air filter must be replaced every 25 hours of functioning or when it result particulary worn.
For replacement, lift the filter and replace with a new one.
Only use original CA-MI filter
DON'T USE THE DEVICE WITHOUT AIR FILTER
The EOLO atomiser does not need maintenance or lubrication.
It is necessary to check functioning and instrument before every use. Unpack the instrument and always
check integrity of plastic parts and feeding cable., they might have been damaged during previous use.
Connect cable to electrical network and turn switch on.
Close the compressor mouthpiece with a finger and verify that loud noises are not present, these can indicate
wrong functioning. Verify that the atomiser is not damaged by previous use (it was badly put away or badly
knocked). A protection fuse (F 1.6 A 250V) not reachable from exterior protects the instrument.
For fuse replacing, please make reference to manufacturer technical personnel.
Fault type
1. Low Nebulization
2. Low Nebulization
Clogged Nebulizer Tank
3. Absence of Nebulization
Clogged Nebulizer Tank
4. Slow Nebulization
5. Noisy Device
Fault 1 - 2 - 3 – 4 – 5
No solution with previous
If the unit does not nebulizer once the above conditions have been checked, we suggest to contact your
dealer or technical service CA-MI.
BEFORE EVERY CHECKING OPERATION, IN CASE OF ANOMALIES OR BAD FUNCTIONING,
PLEASE CONTACT CA-MI TECHNICAL SERVICE
CA-MI DOES NOT GIVE GUARANTEE IF INSTRUMENT, AFTER THE TECHNICL SERVICE
CHECKING, APPEARS TO BE TAMPERED
EOLO est un compresseur à alimentation électrique doté d'un nébuliseur pour les thérapies par
aérosols. L'appareil, conçu pour offrir une facilité de transport et une utilisation presque continue, est
indiqué pour la nébulisation de médicaments bronchodilatateurs et les antibiotiques.
HI-FLO KIT
Air Tube and Mouth-piece)
Air Filter (+n°1 spare part)
MAINTENANCE
Cause
Clogged Nebulizer Tank
Clean and disinfect the nebulizer tank as
If cleaning was not succesful change cruet
medication; Make sure that the nebulizer is
Highly dense drug
Extended use
items
16
Solution
explained in the instruction manual
Check that the nebulizer contains
not clogged;
Check the connection between the
compressor air outlet port and the
accessories
Dilute drug in physiological liquid
Call retainer or manufacturer CA-MI
Call retainer or manufacturer CA-MI